Pulse Dunplicator Tester

About This Product:

1、Model: MPD-1000

2、The product meets the ISO 5840 and GB 12279-2008 cardiovascular implant artificial heart

valve standard、"6.5 pulse flow test" on test equipment Requirements;

3、Arterial pressure: 0 ~ 300mmHg

4、Ventricular pressure: -20 ~ 300mmHg

5、The stroke volume: ≥ 100mL

6、Cardiac output: ≤10L/min

7、Test frequency: 1 ~ 240bpm

8、Suitable valve types: Mechanical valve、bioprosthetic valve、interventional valve、valved conduit

and single mesh

9、Suitable valve diameter: Mechanical valve、bioprosthetic valve: Φ18 ~ Φ34mm

Interventional valve: Φ18 ~ Φ34mm

Valved Piping: Φ11 ~ Φ28mm

Single patch: Φ11 to Φ28mm

10、Test fluid type: physiological saline、glycerin and water ratio、solution or blood

11、The software has sensor calibration、data acquisition、(left heart、right heart) analysis、

calculation、report output、and other functions;

12 、The software has the functions of ventricular waveform generation、waveform downloading、

waveform reading、monitoring and so on、The pressure driven waveform can be adjusted arbitrarily;

13 、Pressure measurement accuracy: ± 1.0mmHg

14、Test liquid temperature range: Room temperature ~ +40℃

15 、Test temperature control accuracy: ±0.5℃

16、The working environment temperature: +15 ~ +35℃、The humidity ≤85%

DEG C without dew

17、Power: 220V、50Hz、400W